× All the transactions performed on 2.0 is available here. To access Market click "Home".

digital NECK vernier calliper (INSIZE)(1187-150)

` 8,062.00

  • MRP: `12000.00

With Thumb Roller The ZERO/ABS key allows the display to be Zero-Set at any slider position along the scale for incremental comparison measurements. This switch will also allow return to the absolute ABS coordinate and display of the true position from the origin point usually jaws-closed point. Large and clear LCD readout...

  • Availability: 50 In Stock
  • Min. Qty. Per Consignee: 2
  • Tape measuring accuracy - + - 0.1 mm
  • Uncertainty over a length of 1000 mm including additional 100 mm on either side of the scale - + - 0.03 mm
  • The instrument should have an appropriate fully automatic mechanism - Yes
  • A graduated linear scale shall be then placed alongside the tape measure. - Yes
  • Each graduation on the tape for every meter will be then calibrated using an appropriate standard - Yes
Other Specifications Download Specification
Type Of Flexible Tape To Be Calibrated Plastic
Tape Measuring Accuracy + - 0.1 mm
Uncertainty Over A Length Of 1000 Mm Including Additional 100 Mm On Either Side Of The Scale + - 0.03 mm
The Instrument Should Have An Appropriate Fully Automatic Mechanism Yes
    One Metre Length Of Tape Is To Be Mounted On The Machine Without Damaging The Tape Or Its Assembly. A Tensile Load Of 50N Or 20N Is To Be Applied To The Tape. Yes
A Graduated Linear Scale Shall Be Then Placed Alongside The Tape Measure. Yes
Every Meter Length Of The Tape Will Be Securely Mounted On A Smooth Surface Yes
A Microscope Assembly Will Be Used To Focus On Both The Tape Measure And The Scale. Yes
Each Graduation On The Tape For Every Meter Will Be Then Calibrated Using An Appropriate Standard Yes
The Microscope Assembly Should Also Allow For The Simultaneous Comparison Of The Tape Measure Against The Linear Scale Mounted Alongside. Yes
The Results Shall Be Stored And Printable In Appropriate Format For Records Yes
To Securely Place The Tape On A Measuring Platform And Apply A Variable Tensile Load Of 50N/20N Using Standard Hexagonal Cast Iron Weight To The Tape And To Maintain It Over The Length Of Tape Being Calibrated Yes
To Allow Mounting Of Tape Measure Of Length Up To 1m And To Enable The Checking Of Tape For Every Meter Length Till Its Total Length On The Testing Platform Alongside The Graduated Linear Scale. The Tape Or Its Assembly Should Not Be Damaged In Any Way During The Whole Process. Yes
Have A Reference/ Standard NPL Calibrated Graduated Metal Linear Scale Of Length 1000 Mm With Accuracy 0.05 Mm And An Additional 100 Mm On Either Side Main Scale Which Can Be Mounted Alongside The Tape Measure For Manual Comparison Using The Microscope Assembly. This Scale Should Be Detachable For Verification Against Secondary Standard Meter Bar. An Appropriate Box To Keep The Linear Scale May Also Be Provided With Each Equipment As Standard. Yes
Have A Suitable Microscope Assembly With An Appropriate Etched Glass Scale Etc. To Focus On Both The Tape As Well As The Linear Scale. Yes
A. The Microscope Assembly Should Have A Lighting System ( And An Additional Battery Powered System, If The Main System Is Connected To Electrical Mains) So That The Area Under Microscope Is Well Lit And Have A Magnification Of Not Less Than 5X Yes
B. The Scale Shall Be Able To Traverse Vertically For A Distance Of 100 Mm Over The Datum Line And Measure To Accuracy: 0.5 mm
C.    The Microscope Assembly Should Also Allow For Manual Comparison Of The Tape Measure Against The Standard Linear Scale With An Accuracy Of ±0.1 Mm And Uncertainty Of 0.03 Mm. Yes
Have A Mechanism To Mount The Tape Measure On A Platform Alongside The Standard Line Scale For Manual Comparison Yes
Length Measuring Instrument Shall Be Capable To Calibrate Or Test Tape As Per OIML R-66:1985(E) For Length Measuring Instrument And Meets The Requirement Of The Same. Yes
Use Calibrated Standard Products For Manufacturing The Tape Measurement Bench I.e. Microscopes, Digital Indicator To Indicate Calibration Data Yes
Length Measuring Instrument Shall Be NPL Calibrated And Complete Type Tested From A National Physical Laboratory With Type Test Report With Their Stamp As It Shall Be Used As Standard Calibrator For Testing Of Tape And Shall Be Provided With Each Equipment As Standard And Also Available With Manufacture For Verification Of The Products. Yes
OIML R-66:1985(E) For Length Measuring Instrument Is For Guidance And All The Gazette Of India Related To Legal Metrology Issued Time To Time , Are Applicable To The Specification As Well As To The Product. Yes
As Per The Gazette Of India Extraordinary, The Legal Metrology Act, 2009,cl. 22,Approval Of Model Before Manufacturing And Cl. 23,Prohibition On Manufacture, Repair Or Sale Without Licence For Product Shall Be Applicable On Manufacture Of Measuring Instrument Yes
Others Certificate Ie ROHS,WHO,GMP Etc. NA
Availability Of Type Test Reports Of Product From Any ILAC/NABL Accredited/Central Gov. LAB. Showing Compliance To The Specification Yes
Test Report And Approvals To Be Furnished To The Buyer On Demand Yes
Test Cetificate No. 1712729A
Teat Report Date 25-04-2011
Name Of The Lab Where Test Conducted. INSIZE
Warranty Time 1 Year
Product Code Available On Supplier Website Yes
If Yes,Product Code No. NA

Price Dekho Price Dekho Price List in india